How to Recalibrate a Temperature Sensor After an Open Circuit

Mar 26, 2026

Leave a message

The question you raised regarding how to recalibrate a temperature sensor following an open circuit addresses a critical step in restoring the accuracy of a temperature measurement system. I fully understand the cautious mindset-the concern that, even after hardware repairs are complete, inaccurate measurements might persist and compromise production quality; indeed, calibration is the core process for re-establishing the system's credibility.

Once a temperature sensor (specifically, a resistance thermometer) that has suffered an open circuit has been repaired, it cannot be immediately put back into service; it must undergo a calibration of its resistance-temperature characteristics. The standard procedure involves placing the repaired sensing element alongside a standard thermometer within a constant-temperature bath. Output resistance values are then compared at multiple temperature points-such as 0°C, 50°C, and 100°C-to calculate any deviations. Based on these findings, the correction coefficients within the display instrument are adjusted, or the sensor's reference table is re-calibrated.

 

I. Prerequisites for Calibration: Ensure Physical Repairs Are Complete

Reliable Open-Circuit Repair: Use arc welding or precision spot welding to restore the connections of the platinum or copper wires, ensuring there are no loose or intermittent connections.

Insulation Restored: Apply high-temperature insulating varnish to the welded joints or encase them in insulating sleeving to prevent short circuits.

Structural Integrity Restored: Rewind and secure the sensing coil to ensure it is firmly seated without any looseness; if necessary, perform a drying treatment (baking at 80–100°C for 2 hours).

The calibration process may only commence once the physical repairs are fully completed and the insulation resistance measures greater than 100 MΩ.

 

II. Standard Calibration Procedure (Using a Pt100 Platinum Resistance Thermometer as an Example)

1. Preparation of Calibration Equipment

Constant-temperature oil bath or water bath (temperature control accuracy: ±0.1°C)

Secondary Standard Platinum Resistance Thermometer (or higher)

High-precision Digital Multimeter (resolution: 0.01 Ω)

Temperature Control and Data Acquisition System (for data recording)

2. Multi-Point Temperature Comparison Method

Align the sensing element of the repaired Pt100 with that of the standard thermometer, and immerse both simultaneously into the constant-temperature bath. Once the temperature has stabilized, record the following data:

Calibration Point

Standard Temperature

Standard Resistance (Ω)

Measured Resistance (Ω)

Deviation (Ω)

Correction Value (°C)

0°C

0.00°C

100.00

100.15

+0.15

-0.38

50°C

50.00°C

119.40

119.62

+0.22

-0.55

100°C

100.00°C

138.51

138.80

+0.29

-0.73

Note: For a Pt100, a change of 1°C corresponds to a resistance change of approximately 0.385 Ω. The deviation is converted into a temperature error using the formula: Temperature Error (°C) = Deviation (Ω) / 0.385.

3. Processing of Calibration Results

If the deviation is linear: Compensation can be applied by setting the "Zero Offset" and "Gain" parameters within the display instrument.

If the deviation is non-linear: It is necessary to re-calibrate the reference table (lookup table) within the control system, or to enable the multi-point non-linear correction function.

If the deviation is excessive (>0.5°C) or unstable: It is recommended to replace the component with a new one, as the repair process may have compromised its long-term stability.

 

III. Simple On-Site Verification (Not a Substitute for Standard Calibration)

If a constant-temperature bath is unavailable, a single-point verification may be performed:

Insert the repaired sensing element into an ice-water mixture (ensuring a temperature of 0°C);

Measure its resistance value; for a Pt100, the reading should fall within the range of 100.0 to 100.3 Ω;

If the reading falls outside this range, the quality of the repair must be re-examined, or replacement of the component should be considered.

This method verifies only the 0°C point; it cannot serve as a substitute for multi-point calibration and is intended solely for preliminary assessment purposes.

info-1328-915

Send Inquiry
Contact usif have any question

You can either contact us via phone, email or online form below. Our specialist will contact you back shortly.

Contact now!